CT3400 Extractive Gas Analyser

CT3400 Extractive Gas Analyser

Cascade’s CT3400 system is designed for the continuous and real time monitoring of multiple gases in industrial processes or laboratories, including engine emissions monitoring, combustion and incineration analysis, among others. Configured as an extractive system, the CT3400 will continuously draw gas through the measurement cell, providing more accurate and sensitive measurements of the gas composition.

 Multicomponent process gas QCL analyser

  • Multicomponent – up to 10 gases
  • Large dynamic range to meet your measurement needs – single ppb to %
  • Fast measurement response - up to 10 Hz for all gases
  • Automated verification reporting for legislated applications

Integrated sample handling system

  • Sintered steel filtration for high particulate gas streams
  • Automated cell and filter purge
  • Gas inlet pressure – up to 2 bar
  • Gas temperatures up to 190 °C

Field serviceable and field configurable

  • Interchangeable modular configuration for up to 4 QCL lasers
  • Low maintenance and low lifetime cost

Typical applications

  • Gas turbine
  • Ammonium nitrate precursors
  • DeNOx / SCR
  • Engine test
  • Laboratory testing

Other applications available on request

  • Value CT3400
    Application Extractive Gas Analyser
    Measurement technique IR Absorption Spectroscopy
    IR source Quantum Cascade Laser [up to 4 lasers]
    Laser classification CLASS 1
      BS EN 60825-1:2007 Safety of laser products
      Equipment classification and requirements
      (identical to IEC 60825-1 2007)
       
    Performance  
    Repeatibility ± 2%
    Accuracy ± 2%
    Linearity R2 > 0.999
    Measurement rate up to 10Hz
       
    Environmental  
    Temperature range 5 to 45 °C
    Sample gas temperature range 20 to +190 °C factory set
    Humidity range 10 to 95 %, non-condensing
    Protection class IP20
    Hazardous area classification n/a
       
    Inputs and Outputs  
    Analogue signal out 6 x 4-20mA measurement output [option]
    Analogue signal in 4-20mA signal input [option]
    Digital signal out 3 x normally closed contact fault relay outputs
    Digital signal out Modbus TCP, 100Base T Ethernet
    Inlet gas port connector 6mm Swagelok type
    Exhaust gas port connector 6mm Swagelok type
       
    Electrical Rating  
    Power supply 120VAC 60Hz / 240VAC 50Hz 200VA
       
    Dimensions  
    Size 759 x 445 x 653mm [HxWxD]
    Weight 83 kg
    Installation Floor standing, castor wheels [option]
  • Name Symbol Units Range LOD
    Nitric oxide NO ppm 5000 0.2
    Nitrogen dioxide NO2 ppm 1000 0.05
    Carbon monoxide CO ppm 10000 0.5
    Carbon dioxide CO2 % 15 0.001
    Sulphur dioxide SO2 ppm 2000 0.5
    Methane CH4 ppm 10000 0.5
    Water H2O % 20 0.005
    Nitrous oxide N2O ppm 5000 0.005
    Ammonia NH3 ppm 2000 0.1
    Formaldehyde HCHO ppm 2000 0.1
    Hydrogen peroxide H2O2 ppm 10,000 0.6
    Ethyne C2H2 ppm 2000 0.2
    Ethene C2H4 ppm 1000 0.05
    Ethane C2H6 ppm 1000 0.05
    Propane C3H8 ppm 2000 0.1
    Butane C4H10 ppm 2000 0.2
    Methanol CH3OH ppm 3000 0.1
    Nitromethane CH3NO2 ppm 2000 0.1

Datasheet

Product Enquiry

 